Resumen
Objectives: The study aims to measure the strength of the proximal fixation of endografts in short and long necks. Design: Three types of endografts were compared: Gore Excluder®, Vascutek Anaconda® and Medtronic Endurant®. Materials and methods: The proximal part of the stent grafts was inserted in bovine arteries and the graft was then attached to a tensile testing machine. The force to obtain dislodgement (DF) from the aorta was recorded for each graft at proximal seal lengths of 10 and 15 mm. Results: The median DF (interquartile range, IQR) for the Excluder, the Anaconda and the Endurant with a seal length of 15 mm was: 11.8 (10.5-12.0) N, 20.8 (18.0-30.1) N and 10.7 (10.4-11.3) N. With the shorter proximal seal of 10 mm, DF was, respectively: 6.0 (4.5-6.6) N, 17.0 (11.2-36.6) N and 6.4 (6.1-12.0) N. Conclusions: The proximal fixation of the Anaconda is superior to the Excluder and the Endurant at short necks of 10 and 15 mm in an experimental set-up. There is a statistically significant decrease of proximal fixation for the Excluder stent graft, when decreasing the length of the proximal neck from 15 to 10 mm.
Idioma original | Inglés |
---|---|
Páginas (desde-hasta) | 187-192 |
Número de páginas | 6 |
Publicación | European Journal of Vascular and Endovascular Surgery |
Volumen | 39 |
N.º | 2 |
DOI | |
Estado | Publicada - feb. 2010 |
Publicado de forma externa | Sí |