T1 - Portable single-lead electrocardiogram device is accurate for QTc evaluation in hospitalized patients
AU - Marín O, María Mónica
AU - García P, Ángel Alberto
AU - Muñoz V, Oscar Mauricio
AU - Castellanos R, Julio César
AU - Cáceres M, Edward
AU - Santacruz P, David
N1 - Publisher Copyright:
© 2021 Heart Rhythm Society
PY - 2021/8
Y1 - 2021/8
N2 - Background: Many commonly used drugs can prolong the QTc interval (QTc), which can lead to potentially life-threatening arrhythmias. In the current era of the COVID-19 pandemic, it is worth mentioning that the disease itself and several drugs used for its treatment have been associated with QTc prolongation. Objective: To evaluate the agreement and clinical precision of a portable single-lead electrocardiogram (ECG) device to measure the QTc interval compared to the standard 12-lead ECG. Methods: In sequential tests, QTc of ECG recordings obtained with the KardiaMobile (KM-1L) device (AliveCor, San Francisco, CA) were compared to QTc obtained with conventional 12-lead ECG. Agreement was evaluated using Bland-Altman plots and Lin's concordance coefficient. Consistency between the 2 devices in determining QTc prolongation (QTc ≥470 ms in males or ≥480 ms in females) was evaluated with kappa statistics. Results: A total of 128 patients with a presumed or confirmed diagnosis of COVID-19 admitted to a university hospital were included. QTc intervals measured with KM-1L were similar to QTc measured with conventional ECG (442.45 ± 40.5 vs 441.65 ± 40.3 ms, P = .15). Bland–Altman analysis showed no significant difference in QTc values (average difference of -0.797, 95% limits of agreement:-13.179; 11.585). Lin's concordance coefficient showed an excellent agreement (0.988, P < .001). Concordance between the 2 devices for determining QTc prolongation was excellent (kappa >0.90). Conclusion: ECG recordings obtained with KM-1L allow an accurate QTc interval assessment. Considering its simplicity of use, this approach has advantages over conventional ECG and can provide an alternative for the evaluation of QTc in hospitalized patients, during the current time of the COVID-19 pandemic and beyond.
AB - Background: Many commonly used drugs can prolong the QTc interval (QTc), which can lead to potentially life-threatening arrhythmias. In the current era of the COVID-19 pandemic, it is worth mentioning that the disease itself and several drugs used for its treatment have been associated with QTc prolongation. Objective: To evaluate the agreement and clinical precision of a portable single-lead electrocardiogram (ECG) device to measure the QTc interval compared to the standard 12-lead ECG. Methods: In sequential tests, QTc of ECG recordings obtained with the KardiaMobile (KM-1L) device (AliveCor, San Francisco, CA) were compared to QTc obtained with conventional 12-lead ECG. Agreement was evaluated using Bland-Altman plots and Lin's concordance coefficient. Consistency between the 2 devices in determining QTc prolongation (QTc ≥470 ms in males or ≥480 ms in females) was evaluated with kappa statistics. Results: A total of 128 patients with a presumed or confirmed diagnosis of COVID-19 admitted to a university hospital were included. QTc intervals measured with KM-1L were similar to QTc measured with conventional ECG (442.45 ± 40.5 vs 441.65 ± 40.3 ms, P = .15). Bland–Altman analysis showed no significant difference in QTc values (average difference of -0.797, 95% limits of agreement:-13.179; 11.585). Lin's concordance coefficient showed an excellent agreement (0.988, P < .001). Concordance between the 2 devices for determining QTc prolongation was excellent (kappa >0.90). Conclusion: ECG recordings obtained with KM-1L allow an accurate QTc interval assessment. Considering its simplicity of use, this approach has advantages over conventional ECG and can provide an alternative for the evaluation of QTc in hospitalized patients, during the current time of the COVID-19 pandemic and beyond.
KW - COVID-19
KW - Electrocardiography
KW - KardiaMobile
KW - Mobile applications
KW - QT/QTc interval
KW - Single-lead ECG device
KW - Smartphone
KW - Ventricular arrhythmias
UR - http://www.scopus.com/inward/record.url?scp=85124147500&partnerID=8YFLogxK
U2 - 10.1016/j.hroo.2021.06.005
DO - 10.1016/j.hroo.2021.06.005
M3 - Article
AN - SCOPUS:85124147500
SN - 2666-5018
VL - 2
SP - 382
EP - 387
JO - Heart Rhythm O2
JF - Heart Rhythm O2
IS - 4
ER -