TY - JOUR
T1 - Diabetic foot prevention
T2 - Repeatability of the loran platform plantar pressure and load distribution measurements in nondiabetic subjects during bipedal standing: A pilot study
AU - Zequera Diaz, Martha Lucia
AU - Garavito Goubert, Leonardo
AU - Sandham, William
AU - Bernal, Juan Camilo
AU - Rodriguez, Ángela
AU - Jiménez, Luis Camilo
AU - Hernández, Andrés
AU - Wilches, Carlos A.
AU - Cecilia Villa, Ana
PY - 2011
Y1 - 2011
N2 - This study was designed to assess the repeatability of the Loran Platform and evaluate the variability of plantar pressure and postural balance, during barefoot standing in nondiabetic subjects, for future diabetic foot clinical evaluation. Measurements were taken for eight nondiabetic subjects (4 females, 4 males, aged 47± 7.2 years) who had no musculoskeletal symptoms. Five variables were measured with the platform in the barefoot standing position. Ten measurements were taken using two different techniques for feet and posture positioning, during three sessions, once a week. For most measurements, no significant effect over time was found with Student's t-test (P<.000125). The ANOVA test of statistical significance confirmed that measurement differences between subjects showed higher variations than measurements taken from the same subject (P<.001). The measurements taken by the Loran Platform system were found to be repeatable.
AB - This study was designed to assess the repeatability of the Loran Platform and evaluate the variability of plantar pressure and postural balance, during barefoot standing in nondiabetic subjects, for future diabetic foot clinical evaluation. Measurements were taken for eight nondiabetic subjects (4 females, 4 males, aged 47± 7.2 years) who had no musculoskeletal symptoms. Five variables were measured with the platform in the barefoot standing position. Ten measurements were taken using two different techniques for feet and posture positioning, during three sessions, once a week. For most measurements, no significant effect over time was found with Student's t-test (P<.000125). The ANOVA test of statistical significance confirmed that measurement differences between subjects showed higher variations than measurements taken from the same subject (P<.001). The measurements taken by the Loran Platform system were found to be repeatable.
UR - http://www.scopus.com/inward/record.url?scp=84856930761&partnerID=8YFLogxK
UR - https://www.hindawi.com/journals/jece/2011/136936/
U2 - 10.1155/2011/136936
DO - 10.1155/2011/136936
M3 - Article
AN - SCOPUS:84856930761
SN - 2090-0147
JO - Journal of Electrical and Computer Engineering
JF - Journal of Electrical and Computer Engineering
M1 - 136936