TY - JOUR
T1 - Novel electronic device to quantify the cyclic fatigue resistance of endodontic reciprocating files after using and sterilization
AU - Zubizarreta-Macho, Álvaro
AU - Alonso-Ezpeleta, Óscar
AU - Martínez, Alberto Albaladejo
AU - Matoses, Vicente Faus
AU - Brucheli, Javier Caviedes
AU - Agustín-Panadero, Rubén
AU - Álvarez, Jesús Mena
AU - Martínez-Berganza, Fernando Vizmanos
N1 - Publisher Copyright:
© 2020 by the authors.
PY - 2020/7
Y1 - 2020/7
N2 - Background: The aim of this study was to analyze the effects of the time of use (TU) and sterilization cycles (SC) of endodontic reciprocating files on cyclic fatigue resistance. Methods: One-hundred-and-twenty (120) Procodile NiTi endodontic reciprocating instruments were selected at random and distributed into the following study groups: A: 0 sterilization cycles/0s time of use (n = 10); B: 0/60 (n = 10); C: 0/120 (n = 10); D: 1/0 (n = 10); E: 1/60 (n = 10); F: 1/120 (n = 10); G: 5/0 (n = 10); H: 5/60 (n = 10); I: 5/120 (n = 10); J: 10/0 (n = 10); K: 10/60 (n = 10); and L: 10/120 (n = 10). A dynamic cyclic fatigue device was designed using computer-aided design/computer-aided engineering (CAD/CAE) technology and created with a 3D printer to simulate the pecking motion performed by the clinician. Failure of the endodontic rotary instrument was detected by a light-emitting diode-light-dependent resistor (LED-LDR) system controlled by an Arduino driver complex and management software. The results were analyzed using the ANOVA test. Results: All pairwise comparisons presented statistically significant differences between the time to failure, number of cycles to failure and number of cycles of in-and-out movement for the time of use study groups (p < 0.001), but not in the number of sterilization cycles (p > 0.05). Conclusions: The time of use of NiTi endodontic reciprocating files negatively affects dynamic cyclic fatigue resistance. Dynamic cyclic resistance is not affected by the number of sterilization cycles.
AB - Background: The aim of this study was to analyze the effects of the time of use (TU) and sterilization cycles (SC) of endodontic reciprocating files on cyclic fatigue resistance. Methods: One-hundred-and-twenty (120) Procodile NiTi endodontic reciprocating instruments were selected at random and distributed into the following study groups: A: 0 sterilization cycles/0s time of use (n = 10); B: 0/60 (n = 10); C: 0/120 (n = 10); D: 1/0 (n = 10); E: 1/60 (n = 10); F: 1/120 (n = 10); G: 5/0 (n = 10); H: 5/60 (n = 10); I: 5/120 (n = 10); J: 10/0 (n = 10); K: 10/60 (n = 10); and L: 10/120 (n = 10). A dynamic cyclic fatigue device was designed using computer-aided design/computer-aided engineering (CAD/CAE) technology and created with a 3D printer to simulate the pecking motion performed by the clinician. Failure of the endodontic rotary instrument was detected by a light-emitting diode-light-dependent resistor (LED-LDR) system controlled by an Arduino driver complex and management software. The results were analyzed using the ANOVA test. Results: All pairwise comparisons presented statistically significant differences between the time to failure, number of cycles to failure and number of cycles of in-and-out movement for the time of use study groups (p < 0.001), but not in the number of sterilization cycles (p > 0.05). Conclusions: The time of use of NiTi endodontic reciprocating files negatively affects dynamic cyclic fatigue resistance. Dynamic cyclic resistance is not affected by the number of sterilization cycles.
KW - Cyclic fatigue
KW - Endodontics
KW - Reciprocating movement
KW - Sterilization cycles
KW - Time of use
UR - http://www.scopus.com/inward/record.url?scp=85088636124&partnerID=8YFLogxK
U2 - 10.3390/app10144962
DO - 10.3390/app10144962
M3 - Article
AN - SCOPUS:85088636124
SN - 2076-3417
VL - 10
JO - Applied Sciences (Switzerland)
JF - Applied Sciences (Switzerland)
IS - 14
M1 - 4962
ER -