Skip to main navigation Skip to search Skip to main content

Explantador de Cabrera: un nuevo dispositivo para explantar endoprótesis aórticas después de reparo endovascular de aneurisma de aorta, un estudio experimental ex vivo

Translated title of the contribution: Cabrera explanter: a new device to explant aortic stents after endovascular aortic aneurysm repair, an ex vivo experimental study
  • Luis Felipe Cabrera-Vargas
  • , Iván Lozada-Martínez
  • , Nicolás Forero-Ramírez
  • , Hernando Thorne-Vélez

Research output: Contribution to journalArticlepeer-review

Abstract

Introduction. Complications after endovascular aneurysm repair (EVAR) can be resolved with endovascular techniques; however, when indicated, stent explantation is a complex procedure, which is associated with vascular or visceral injuries, with high morbidity and mortality in patients, with advanced age and multiple comorbidities, and therefore high surgical risk. There are no devices produced by the industry to explant aortic endoprostheses, so the objective of this work was to develop a device for the explantation of aortic endoprostheses. Methods. An experimental study was carried out, in the preclinical phase, to develop a device for the explantation of aortic endoprostheses, with tests in 3D models and in a cadaveric porcine animal model. Results. It is feasible to develop an experimental model of a new device for explanting aortic endoprostheses, called Cabrera explanter, and verify its operation in a cadaveric animal model. The use of the Cabrera explanter limited damage to the aortic wall by the endoprosthesis by 100% at the time of explantation in an ex vivo experimental model. Conclusions. Using a septum syringe, the Cabrera explanter is superior to the standard stent explantation technique by limiting injury to the aortic wall, collapsing and releasing the adrenal fixation hooks in a controlled and safe manner into the aortic lumen, and subsequently, extract it quickly and effectively, preserving the greatest amount of healthy aorta for the subsequent aorto-iliac reconstruction.

Translated title of the contributionCabrera explanter: a new device to explant aortic stents after endovascular aortic aneurysm repair, an ex vivo experimental study
Original languageSpanish
Pages (from-to)580-589
Number of pages10
JournalRevista Colombiana de Cirugia
Volume39
Issue number4
DOIs
StatePublished - 05 Jul 2024
Externally publishedYes

Fingerprint

Dive into the research topics of 'Cabrera explanter: a new device to explant aortic stents after endovascular aortic aneurysm repair, an ex vivo experimental study'. Together they form a unique fingerprint.

Cite this